Cytofluor 4000: Difference between revisions
Jump to navigation
Jump to search
| Line 1: | Line 1: | ||
=== Fluorescence Multi-Well Plate Reader === | ===Fluorescence Multi-Well Plate Reader=== | ||
{| class="wikitable" | {| class="wikitable" | ||
|+CytoFluor command list | |+CytoFluor command list | ||
| Line 89: | Line 87: | ||
| | | | ||
|} | |} | ||
==== List of major IC ==== | |||
* P80C32SPBN | |||
* MAX699 | |||
* 24LC01B | |||
* MAX232CPE | |||
* 74AC273PC | |||
* MC145053P | |||
* CY545 | |||
* PALCE22V10Q-25PC/4 | |||
* PSD311-B-15 | |||
Revision as of 02:58, 21 June 2025
Fluorescence Multi-Well Plate Reader
| command | parameters | function | |
|---|---|---|---|
| RS | reset? | ||
| C | Center (seems same as RS) | ||
| EM | ### | Position Emission Filter | (for loading) RS\r EM 167\r |
| EX | ### | Position Excitation filter | |
| L | 0|1 | Lamp control 0=ON 1=OFF | L 0\r |
| O | Plate out (eject) | O\r | |
| SA | ### ### | Moves plate [F/Back] [L/R] | Plate in: C\r SA 315 1800\r |
| P | ? | ||
| I | ### | moves plate fwd/back. Light On | ###'s don't seem to do anything |
| GL | Returns a number ### | ||
| GE | ## | Gets value | Returns: Value at {param} is ####### |
| R | Read? | ||
| RP | Returns a set ### #### | ||
| M | ## | Shake fwd/bk for ~#'s seconds | |
| G | ## | ? | |
List of major IC
- P80C32SPBN
- MAX699
- 24LC01B
- MAX232CPE
- 74AC273PC
- MC145053P
- CY545
- PALCE22V10Q-25PC/4
- PSD311-B-15