JEOL JSM-6060: Difference between revisions

From RECESSIM, A Reverse Engineering Community
Jump to navigation Jump to search
 
(One intermediate revision by the same user not shown)
Line 15: Line 15:
Software:
Software:


* MS Windows 32 bit (Windows 7 Pro tested)
* MS Windows 32 bit (Windows 7 Pro tested) (https://archive.org/download/windows-7-iso-)
* MS Windows update package since online support no longer exists
* MS Windows update package since online support no longer exists (https://www.majorgeeks.com/mg/getmirror/simplix_updatepack,1.html)
* Virtual Floppy Drive (https://vfd.sourceforge.net/)
* SCSI Driver
* SCSI Driver


Line 49: Line 50:


==== September 24, 2026 ====
==== September 24, 2026 ====
Downloaded and installed Windows 7 Pro 32 bit.
 
* Downloaded and installed Windows 7 Pro 32 bit.  
* Installed update package, .NET 4.7 and Virtual Floppy Drive.
** VFD is used to load a backup floppy containing the JEOL 6060 backup config files (not original to this machine) during JEOL software installation. Not sure if this step is necessary or if those files can be loaded later.
* Installed JEOL SEM software
** No SCSI or capture cards in the machine so the software hangs upon start and eventually exits

Latest revision as of 15:55, 24 September 2026

Installation and Configuration

JSM-6060

Information on troubleshooting and setting up the SEM hardware

Control Computer

Information on setting up the computer used to run the JEOL software

System Requirements

Hardware:

  • PC with PCI slot(s) (>4GB RAM uncecessary)
  • AHA-2940 SCSI card to communicate with SEM

Software:

Deben Chamberscope

This SEM came with a Deben Chamberscope. This is a video camera placed in the SEM chamber and an external controller that allows you to view the sample while the SEM is under vacuum and operating. The control box contains three video inputs, two DB9 and one BNC.

It is unknown at this time if the SEM electron beam preview image was routed into this device and used by the JEOL software, or if this is purely for viewing on an external monitor.

Roughing Vacuum Pump

The vacuum pump that came with this machine is an ULVAC Model G-100DB. It was leaking oil from the front seal which had to be replaced prior to use. The following parts were purchased.

  • Front outer seal
  • Replacement bearing
  • Front inner seal

Chiller

A water chiller is needed to cool the oil diffusion pump used in the machine. In this JSM-6060 this is the only component that is water cooled.

Voltage Converter

The JSM-6060 was developed by a Japanese company and runs on the Japanese standard 100V. A converter was purchased to transform the 120V US standard to 100V. The machine can run on 50/60Hz so no additional power conditioning is needed.

Cleaning Supplies

This JSM-6060 was purchased from a surplus company and was very dirty. Even a "clean" SEM may not be clean enough for operation. The following supplies were used to clean the system.

  • Everclear (yes, that one)
  • Texwipe TX1036

Testing and Work Log

September 23, 2026

Picked up a couple of computers donated to me by the DFW hacking community to use as the SEM control computer. Installed Windows 7 Pro 64 bit (an EPIC pain) to discover in the end that only 32 bit Windows will work!

September 24, 2026

  • Downloaded and installed Windows 7 Pro 32 bit.
  • Installed update package, .NET 4.7 and Virtual Floppy Drive.
    • VFD is used to load a backup floppy containing the JEOL 6060 backup config files (not original to this machine) during JEOL software installation. Not sure if this step is necessary or if those files can be loaded later.
  • Installed JEOL SEM software
    • No SCSI or capture cards in the machine so the software hangs upon start and eventually exits