JEOL JSM-6060: Difference between revisions
Created page with "== Installation and Configuration == === JSM-6060 === Information on troubleshooting and setting up the SEM hardware === Control Computer === Information on setting up the computer used to run the JEOL software ==== System Requirements ==== Hardware: * PC with PCI slot(s) (>4GB RAM uncecessary) * AHA-2940 SCSI card to communicate with SEM Software: * MS Windows 32 bit (Windows 7 Pro tested) * MS Windows update package since online support no longer exists * SCSI D..." |
|||
| Line 18: | Line 18: | ||
* MS Windows update package since online support no longer exists | * MS Windows update package since online support no longer exists | ||
* SCSI Driver | * SCSI Driver | ||
=== Roughing Vacuum Pump === | |||
The vacuum pump that came with this machine is an ULVAC Model G-100DB. It was leaking oil from the front seal which had to be replaced prior to use. The following parts were purchased. | |||
* Front outer seal | |||
* Replacement bearing | |||
* Front inner seal | |||
=== Chiller === | |||
A water chiller is needed to cool the oil diffusion pump used in the machine. In this JSM-6060 this is the only component that is water cooled. | |||
=== Voltage Converter === | |||
The JSM-6060 was developed by a Japanese company and runs on the Japanese standard 100V. A converter was purchased to transform the 120V US standard to 100V. The machine can run on 50/60Hz so no additional power conditioning is needed. | |||
=== Cleaning Supplies === | |||
This JSM-6060 was purchased from a surplus company and was very dirty. Even a "clean" SEM may not be clean enough for operation. The following supplies were used to clean the system. | |||
* Everclear (yes, that one) | |||
* Texwipe TX1036 | |||
Revision as of 14:20, 24 September 2026
Installation and Configuration
JSM-6060
Information on troubleshooting and setting up the SEM hardware
Control Computer
Information on setting up the computer used to run the JEOL software
System Requirements
Hardware:
- PC with PCI slot(s) (>4GB RAM uncecessary)
- AHA-2940 SCSI card to communicate with SEM
Software:
- MS Windows 32 bit (Windows 7 Pro tested)
- MS Windows update package since online support no longer exists
- SCSI Driver
Roughing Vacuum Pump
The vacuum pump that came with this machine is an ULVAC Model G-100DB. It was leaking oil from the front seal which had to be replaced prior to use. The following parts were purchased.
- Front outer seal
- Replacement bearing
- Front inner seal
Chiller
A water chiller is needed to cool the oil diffusion pump used in the machine. In this JSM-6060 this is the only component that is water cooled.
Voltage Converter
The JSM-6060 was developed by a Japanese company and runs on the Japanese standard 100V. A converter was purchased to transform the 120V US standard to 100V. The machine can run on 50/60Hz so no additional power conditioning is needed.
Cleaning Supplies
This JSM-6060 was purchased from a surplus company and was very dirty. Even a "clean" SEM may not be clean enough for operation. The following supplies were used to clean the system.
- Everclear (yes, that one)
- Texwipe TX1036